Debian Codegraph @

eyes17/helpFiles/hi/_sources/4.2.rst

raw
1 XY plotting
2 ===========
3
4 Oscilloscopes generally plots time X-axis. They also provide and X-Y mode, where one input is plotted on the X-axis. This feature is used for showing lissajou's figures and also for measuring the phase difference of two input waveforms.
5
6 Objective
7 ---------
8
9 Plot two signals in the X-Y mode and find out their phase difference. The phase difference is be generated by CR circuit.
10
11 .. image:: schematics/RCsteadystate.svg
12 :width: 300px
13 .. image:: pics/phase-from-xy-screen.png
14 :width: 300px
15
16 Procedure
17 ---------
18
19 - Make the connections as shown in the figure. Use :math:`1 k\Omega` and :math:`1 \mu F`
20 - Note down the maximum value :math:`y_{max}` and the y-intercept :math:`y_{int}`
21
22 The phase difference is given by the equation :math:`\theta = sin^{-1}\Bigg(\frac{y_{int}}{y_{max}}\Bigg)`
23
24 .. image:: pics/xyplot-screen.png
25 :width: 400px
26
27
28 Discussion
29 ----------
30
31 There is an option to plot (A1-A2) against A2. In the above circuit, this plots the voltage across the
32 capacitor against the voltage across the resistor.
33
34 - Select :math:`C = 1\mu F`, :math:`R = 1 k\Omega` and plot (A1-A2) vs A2.
35 - Adjust the frequency to make both voltages equal.