| 1 |
Temperature measurement using PT100 |
| 2 |
=================================== |
| 3 |
|
| 4 |
The Resistance Temperature Detector (RTD) is a temperature sensing device whose resistance changes with temperature, typically built from platinum. To measure the resistance across an RTD, apply a constant current, measure the resulting voltage, and determine the RTD resistance. The PT100 is a popular RTD with a resistance of :math:`100 \Omega` at 0°C. Resistance of a PT100 element is related to the temperature |
| 5 |
by the equation :math:`R(T) = R_0 (1 + AT + BT^2)`, where |
| 6 |
:math:`A = 3,9083 \times 10^{ 3}` and :math:`B = 5,775 \times 10^{ 7}`. |
| 7 |
|
| 8 |
Objective |
| 9 |
--------- |
| 10 |
|
| 11 |
Record the temperature of a liquid by using a Pt100. |
| 12 |
|
| 13 |
Procedure |
| 14 |
--------- |
| 15 |
|
| 16 |
.. image:: schematics/pt100.svg |
| 17 |
:width: 300px |
| 18 |
|
| 19 |
- Make the connections as shown in the figure. |
| 20 |
- Enter the Gain, Offset error and the Current from CCS |
| 21 |
- Select the temperature range and time intervals |
| 22 |
- Select the required parameters and press START |
| 23 |
|
| 24 |
|
| 25 |
Cooling curve of water is shown in the figure below. |
| 26 |
|
| 27 |
.. image:: pics/pt100-screen.png |
| 28 |
:width: 300px |
| 29 |
|
| 30 |
Discussion |
| 31 |
---------- |
| 32 |
|
| 33 |
The accuracy of the measurements can be increased by the following steps. |
| 34 |
|
| 35 |
Measure the actual current from CCS and enter it in the GUI, it could be slightly |
| 36 |
different from 1.1mA. The input to A3 is amplified 11 times by |
| 37 |
connecting :math:`1~k\Omega` resistor from Rg to Ground. The gain and offset should |
| 38 |
be measured an entered in to the program. |