Debian Codegraph @

eyes17/helpFiles/hi/_sources/1.2.rst

raw
1 The equipment
2 =============
3
4 ExpEYES-17 is interfaced and powered by the USB port of the computer,
5 and it is programmable in Python. It can function as a low frequency
6 oscilloscope, function generator, programmable voltage source, frequency
7 counter and data logger. For connecting external signals, it has connectors
8 on the top panel. as shown in the figure below. The software can monitor and
9 control the voltages at these terminals. Inorder to measure other
10 parameters (like temperature, pressure etc.), we
11 need to convert them in to electrical signals by using appropriate
12 sensor elements. The accuracy of the voltage measurements is decided by
13 the stability of the 3.3V reference used, it is 50ppm per degree
14 celcius. The gain and offset errors are eliminated by initial
15 calibration, using a 16bit ADC. Even though our primary objective is to
16 do experiments, you are advised to read through the brief description of
17 the equipment given below. The device can be also used as a test
18 equipment for electrical and electronics engineering experiments.
19
20 **IMPORTANT :**
21
22 *The external voltages connected to ExpEYES17 must be within the allowed
23 limits. Inputs A1 and A2 must be within ±16 volts range and Inputs IN1
24 and IN2 must be in 0 to 3.3V range. Exceeding these limits may result in
25 damage to the equipment. To measure higher voltages, scale them down
26 using resistive potential divider networks.*
27
28 .. image:: pics/eyes-box.jpg
29 :width: 400px
30
31
32 External connections
33 --------------------
34
35 The functions of the external connections briefly explained below. All
36 the black coulored terminals are at ground potential, all other voltages
37 are measured with respect to it.
38
39 Outputs:
40 ^^^^^^^^
41
42 Constant Current Source (CCS) :
43 """""""""""""""""""""""""""""""
44
45 The constant current source can be switched ON and OFF under software
46 control. The nominal value is 1.1 mA but may vary from unit to unit, due
47 to component tolerances. To measure the exact value, connect an ammeter
48 from CCS to GND. Another method is to connect a known resistance (~1k)
49 and measure the voltage drop across it. The load resistor should be less
50 than 3k for this current source.
51
52 Programmable Voltage (PV1) :
53 """"""""""""""""""""""""""""
54
55 Can be set, from software, to any value in the -5V to +5V range. The
56 resolution is 12 bits, implies a minimum voltage step of around 2.5
57 millivolts.
58
59 Programmable Voltage (PV2) :
60 """"""""""""""""""""""""""""
61
62 Can be set, from software, to any value in the -3.3V to +3.3V range. The
63 resolution is 12 bits.
64
65 Square Wave SQ1:
66 """"""""""""""""
67
68 Output swings from 0 to 5 volts and frequency can be varied 4Hz to
69 100kHz. All intermediate values of frequency are not possible. The duty
70 cycle of the output is programmable. Setting frequency to 0Hz will make
71 the output HIGH and setting it to :math:`-1` will make it LOW, in both cases
72 the wave generation is disabled. SQR1 output has a :math:`100~\Omega` **series
73 resistor** inside so that it can drive LEDs directly.
74
75 Square Wave SQ2:
76 """"""""""""""""
77
78 Output swings from 0 to 5 volts and frequency can be varied 4Hz to
79 100kHz. All intermediate values of frequency are not possible. The duty
80 cycle of the output is programmable. SQR2 is not available when WG is
81 active.
82
83 Digital Output (OD1) :
84 """"""""""""""""""""""
85
86 The voltage at OD1 can be set to 0 or 5 volts, using software.
87
88 Sine/Triangular Wave WG:
89 """"""""""""""""""""""""
90
91 Frequency can be varied from 5Hz to 5kHz. The peak value of the
92 amplitude can be set to 3 volts, 1.0 volt or 80 mV. Shape of the output
93 waveform is programmable. Using the GUI sine or triangular can be
94 selected. WG bar is inverted WG.
95
96 Inputs:
97 ^^^^^^^
98
99 Capacitance meter IN1:
100 """"""""""""""""""""""
101
102 Capacitance connected between IN1 and Ground can be measured. It works
103 better for lower capacitance values, upto 10 nanoFarads, results may not
104 be very accurate beyond that.
105
106 Frequency Counter IN2:
107 """"""""""""""""""""""
108
109 Capable of measuring frequencies upto several MHz.
110
111 Resistive Sensor Input (SEN):
112 """""""""""""""""""""""""""""
113
114 This is mainly meant for sensors like Light Dependent Resistor,
115 Thermistor, Photo-transistor etc. SEN is internally connected to 3.3
116 volts through a 5.1\ k\ Ω resistor.
117
118 :math:`\pm16\ V` Analog Inputs, A1 & A2:
119 """"""""""""""""""""""""""""""""""""""""
120
121 Can measure voltage within the ±16 volts range. The input voltage range
122 can be selected from .5V to 16V fullscale. Voltage at these terminals
123 can be displayed as a function of time, giving the functionality of a
124 low frequency oscilloscope. The maximum sampling rate is 1 Msps
125 /channel. Both have an input impedance of 1\ M\ Ω .
126
127 :math:`\pm3.3\ V` Analog Input A3:
128 """"""""""""""""""""""""""""""""""
129
130 Can measure voltage within the ±3.3 volts range. The input can be
131 amplified by connecting a resistor from Rg to Ground,
132 :math:`gain=1 + \frac{R_g}{10000}`.
133 This enables displaying very small amplitude signals.
134 The input impedance of A3 is 10\ M\ Ω.
135
136 Microphone input MIC:
137 """""""""""""""""""""
138
139 A condenser microphone can be connected to this terminal and the output
140 can be captured.
141
142 I2C Sensor Interface:
143 ^^^^^^^^^^^^^^^^^^^^^
144
145 The four connections (+5V, Ground, SCL and SDA) of the 8 terminal berg
146 strip supports I2C sensors. The software is capable of recognizing a
147 large number of commercially available I2C sensors.
148
149 :math:`\pm\ 6\ V/10\ mA` Power supply:
150 ^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^
151
152 The VR+ and VR- are regulated power outputs. They can supply very little
153 current, but good enough to power an Op-Amp.
154
155 1.2.2 Accessory Set
156 -------------------
157
158 .. image:: pics/accessory.jpg
159 :width: 400px
160
161 To get started, a set of accessories are provided with expEYES.
162
163 - Pieces of wires, with pin and with crocodile clip.
164 - Condenser microphone with leads.
165 - Inductor Coil (2) : 44SWG wire on 1cm dia bobbin. Around 3000 Turns
166 (some may have more turns). These coils can be used for studying
167 inductance, electromagnetic induction etc.
168 - Piezo Electric Discs (2) : Resonant frequency is around 3500 Hz. Can
169 be energized by WG output or SQR1. Discs are enclosed in a plastic
170 shell that forms a cavity, that enhances the amplitude of sound
171 produced.
172 - DC Motor : Should be powered by a DC voltage less than 3 volts.
173 - Permanent Magnets : (a) 10mm dia & length (b) 5 mm dia & 10 mm length (c)
174 Button size magnets(2)
175 - 5mm LEDS : RED, BLUE, GREEN, WHITE
176 - Capacitors : 100pF, 0.1 µF , 1 µF et 22 µF
177 - Inductor : 10 mH / 20Ω,
178 - Resistors : 560Ω, 1k\ Ω, 2.2k\ Ω , 10k\ Ω , 51k\ Ω and 100 k\ Ω
179 - LDR