Debian Codegraph @

eyesjunior/helpFiles/en/_sources/2.2.rst

raw
1 Measuring Resistance
2 ====================
3
4 Resistance is generally measured using a multimeter. ExpEYES provides a limited facility
5 to do this, using the SEN terminal. SEN is actually a voltage measurement terminal, with
6 a :math:`5.1k~\Omega` resistor connected to :math:`3.3V` internally. The internal :math:`5.1k~\Omega`
7 and the external resistor connected to SEN forms a series combination. By knowing the voltage across both
8 and the value of one resistance, we can calculate the other using Ohm's law. The drawback of this method
9 is that the accuracy is poor when the external resistance differs widely from :math:`5.1k~\Omega`.
10 A resonable range is from :math:`100~\Omega` to :math:`100~k\Omega`.
11
12 Objective
13 ---------
14
15 Measure resistance using the **SEN** terminal.
16
17 .. image:: schematics/res-measure.svg
18 :width: 300px
19
20 Procedure
21 ---------
22
23 - Connect the resistor between SEN and any Ground terminal.
24 - Observe the value shown on the top right side of the GUI.
25 - Repeat with different resistors