| 1 |
<h2 id="schematic">Schematic</h2> |
| 2 |
<hr> |
| 3 |
<p><img src="images/pt100.png" alt=""></p> |
| 4 |
<p>PT100 is an RTD sensor, it's resistance changes with temperature. It is 100 Ohms at zero degree celcius. |
| 5 |
The resistance is measured by connecting PT100 to a Constant Current Source. The voltage drop is proportional to the resistance. |
| 6 |
It is measured after some amplification, since the voltage is less than200 mV even at 100 degC. |
| 7 |
</p> |
| 8 |
<hr> |
| 9 |
<ul> |
| 10 |
<li>Connect the PT100 sensor from the CCS to Ground. CCS = 1.1mA nominal</li> |
| 11 |
<li>Connect CCS to A3, for monitoring</li> |
| 12 |
<li>Connect a gain resistor Rg=1K, Gain = 1 + 10k/Rg = 1 + 10/1 = 11</li> |
| 13 |
<li>Select the Total Time and time between measurements</li> |
| 14 |
<li>Select the Minimum and Maximum limits for Temperatures to be measured</li> |
| 15 |
</ul> |
| 16 |
|
| 17 |
<hr> |
| 18 |
<p>For better accuracy, calibrate CCS value, gain and Offset of A3. </p> |
| 19 |
<ul> |
| 20 |
<li>Connect a 1k resistor from CCS to Ground. Measure the voltage to calculate the exact value of CCS</li> |
| 21 |
<li>Connect Rg and measure A3, to get the Offset error at that gain </li> |
| 22 |
<li>Connect PT100. Meaasure A3 with and without Rg, to calculate the gain on A3 </li> |
| 23 |
<li>Enter the calculated values, to get better accuracy</li> |
| 24 |
</ul> |