Debian Codegraph @

docs/source/tutorial/s3_normalization.rst

raw
1 Setting normalization and other options
2 ^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^
3 .. _s3p4:
4
5 This section teaches you how to set up normalization by a beamstop counter, and
6 other options. It assumes you have completed :ref:`Parts 1 <s3p3>` and :ref:`2 <s3p1>`\ .
7
8 A video version of this tutorial is available:
9
10 .. raw:: html
11
12 <style>.embed-container { position: relative; padding-bottom: 56.25%; height: 0; overflow: hidden; max-width: 100%; } .embed-container iframe, .embed-container object, .embed-container embed { position: absolute; top: 0; left: 0; width: 100%; height: 100%; }</style><div class='embed-container'><iframe src='https://www.youtube.com/embed/NTjrds0stGc' frameborder='0' allowfullscreen></iframe></div>
13
14 The written version of the tutorial follows.
15
16 #. Open the Options window by selecting “Advanced Options” in the Options menu.
17
18 #. In the window that shows up select the Image/Header Format section on the left.
19 In the area on the right click the Load Image button.
20
21 |config_norm_load_image_png|
22
23 #. In the window that pops up, select the **agbe_008_0001.tif** file. Click
24 the Open button.
25
26 * *Note:* You can select any image of the appropriate type, not just the behenate.
27
28 #. In the Image/Header Format window you should now see header values loaded into the
29 list. Click the Apply button at the bottom of the screen.
30
31 |config_norm_apply_png|
32
33 #. Click on the Normalization section in the options list on the left.
34
35 #. In the fields at the bottom of the Normalization panel, make sure “/” is selected
36 in the left drop-down menu, and enter I1 in the large field.
37
38 * *Note:* It is typical in SAXS to normalize by the transmitted intensity. At the
39 BioCAT beamline, the beamstop counter is name I1, which is why we are using
40 that name in the normalization expression.
41
42 * *Tip:* You can use the large field as a drop-down menu to see and select
43 available normalization counters.
44
45 #. Click the Calc button to evaluate the expression for the counter values loaded
46 in the Image/Header Format tab. You should get a value of 7200.0.
47
48 #. Click the Add button to add the expression to the normalization list.
49
50 |config_norm_values_png|
51
52 #. You should now see the normalization in the normalization list.
53 Make sure the “Enable Normalization” checkbox at the top of the page is checked.
54
55 |config_norm_enable_png|
56
57 #. Click OK to exit the options window.
58
59 #. In the file list, select the **agbe_008_0001.tif** file and click the Plot
60 button. You will see a curve get plotted in the top panel of the Profiles plot.
61
62 #. Click on the Profiles tab. You will see a profile loaded in the Profiles list.
63
64 |config_norm_start_point1_png|
65
66 #. Adjust the start point for q Min to remove the points with zero value at the start of
67 the curve (these are q points entirely in the mask). Set q Min so that the first point
68 is the peak of the curve on the main plot. This should be around point 13 (depending
69 on your mask).
70
71 * *Tip:* It is easier to see the start point if you put the plot on
72 a log-log scale.
73
74 |config_norm_start_point2_png|
75
76 #. Open the Options window as in Step 1.
77
78 #. Click on the Radial Averaging section in the options list on the left. Set “Start plots
79 at q-point number” to the number you just found in Step 13.
80
81 * *Note:* This makes it so that every curve loaded from now on will by default
82 not display the first n points, which are covered by the beamstop.
83
84 * *Tip:* You can do the same for the end point.
85
86 |config_norm_start_point3_png|
87
88 #. RAW also allows you to add arbitrary metadata to your radially integrated
89 files. One use case is to provide metadata keys for data deposition (e.g. in
90 the `SASBDB <https://www.sasbdb.org/>`_). To start, click on the Metadata
91 section in the options list on the left of the Options window.
92
93 #. In the 'Key' field enter 'Detector'. In the 'Value' field enter 'Pilatus3 X 1M'.
94
95 #. Click the Add button.
96
97 |config_norm_metadata1_png|
98
99 #. Click the OK button to exit the options window and save your changes.
100
101 #. You have configured everything necessary, and are now ready to save your settings.
102 Go to the File menu and select “Save Settings”.
103
104 #. Save the settings as **SAXS.cfg**\ .
105
106 #. These settings can now be used to process images, and can be reloaded when you
107 open RAW by selecting “Load Settings” from the File menu.
108
109 **Additional Tips**
110
111 There are other settings you may find it useful to set.
112
113 #. In the "General Settings" section you can set "Hide controls on
114 profile items for new plots". This means that when you load a profile
115 item it starts with the extra controls (such as q min and q max) hidden.
116
117
118
119 .. |config_norm_load_image_png| image:: images/config_norm_load_image.png
120 :target: ../_images/config_norm_load_image.png
121
122 .. |config_norm_apply_png| image:: images/config_norm_apply.png
123 :target: ../_images/config_norm_apply.png
124
125 .. |config_norm_values_png| image:: images/config_norm_values.png
126 :target: ../_images/config_norm_values.png
127
128 .. |config_norm_enable_png| image:: images/config_norm_enable.png
129 :target: ../_images/config_norm_enable.png
130
131 .. |config_norm_start_point1_png| image:: images/config_norm_start_point1.png
132 :target: ../_images/config_start_point1.png
133
134 .. |config_norm_start_point2_png| image:: images/config_norm_start_point2.png
135 :target: ../_images/config_start_point2.png
136
137 .. |config_norm_start_point3_png| image:: images/config_norm_start_point3.png
138 :target: ../_images/config_start_point3.png