| 1 |
Molecular weight analysis |
| 2 |
^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^^ |
| 3 |
.. _raw_mw: |
| 4 |
|
| 5 |
This tutorial covers how to use RAW for molecular weight analysis. This is not a |
| 6 |
tutorial on basic principles and best practices for molecular weight analysis. |
| 7 |
For that, please see the :ref:`SAXS tutorial <saxs_mw>`. |
| 8 |
|
| 9 |
RAW provides four methods of molecular weight analysis: |
| 10 |
|
| 11 |
* Referencing I(0) to that of a known standard |
| 12 |
* From the volume of correlation using the method of Rambo and Tainer |
| 13 |
* From the adjusted Porod volume using the method of Fisher et al. |
| 14 |
* From the value of I(0) on an absolute scale. |
| 15 |
|
| 16 |
If ATSAS is installed, RAW also provides an additional two methods of MW |
| 17 |
calculation using the ATSAS tools, for a total of six different methods: |
| 18 |
|
| 19 |
* From classification by machine learning (ATSAS datclass/Shape&Size) |
| 20 |
* From a Bayesian estimation based on concentration independent methods (ATSAS |
| 21 |
datmw bayes) |
| 22 |
|
| 23 |
If you use these MW methods in RAW, in addition to the RAW paper please cite |
| 24 |
these papers: |
| 25 |
|
| 26 |
* Volume of correlation: Rambo, R.P. & Tainer, J.A. Nature (2013). 496, |
| 27 |
477-481. DOI: `10.1038/nature12070 <https://doi.org/10.1038/nature12070>`_ |
| 28 |
|
| 29 |
* Corrected Porod volume: V. Piiadov, E. Ares de Araujo, M. Oliveira Neto, A. F. |
| 30 |
Craievich, and I. Polikarpov. Protein Science (2019). 28(2), 454-473. |
| 31 |
DOI: `10.1002/pro.3528 <https://doi.org/10.1002/pro.3528>`_ |
| 32 |
|
| 33 |
* Bayesian inference (ATSAS): Hajizadeh, N. R., Franke, D., Jeffries, C. |
| 34 |
M. & Svergun, D. I. (2018). Sci. Rep. 8, 7204. |
| 35 |
DOI: `10.1038/s41598-018-25355-2 <https://doi.org/10.1038/s41598-018-25355-2>`_ |
| 36 |
|
| 37 |
* Shape&Size (ATSAS): Franke, D., Jeffries, C. M. & Svergun, D. I. (2018). |
| 38 |
Biophys. J. 114, 2485–2492. |
| 39 |
DOI: `10.1016/j.bpj.2018.04.018 <https://doi.org/10.1016/j.bpj.2018.04.018>`_ |
| 40 |
|
| 41 |
A video version of this tutorial is available: |
| 42 |
|
| 43 |
.. raw:: html |
| 44 |
|
| 45 |
<style>.embed-container { position: relative; padding-bottom: 56.25%; height: 0; overflow: hidden; max-width: 100%; } .embed-container iframe, .embed-container object, .embed-container embed { position: absolute; top: 0; left: 0; width: 100%; height: 100%; }</style><div class='embed-container'><iframe src='https://www.youtube.com/embed/qYWzhBdnsRY' frameborder='0' allowfullscreen></iframe></div> |
| 46 |
|
| 47 |
The written version of the tutorial follows. |
| 48 |
|
| 49 |
#. In RAW, right click on the subtracted GI scattering profile in the Profiles panel |
| 50 |
and select “Molecular weight.” Alternatively click on the "Mol. Weight" button |
| 51 |
at the bottom of the Profiles panel. |
| 52 |
|
| 53 |
|mw_open_png| |
| 54 |
|
| 55 |
#. At the top of the panel are the results of the Guinier fit. All methods |
| 56 |
require a good Guinier fit, so you can use that button to redo the fit if |
| 57 |
necessary. In the lower part of the panel, the results of the estimates |
| 58 |
for MW are shown. |
| 59 |
|
| 60 |
* *Note:* Neither the I(0) Ref. MW panel nor the Abs. MW panel should be reporting a MW. |
| 61 |
|
| 62 |
* *Note:* If you have ATSAS installed and accessible to RAW, you should see |
| 63 |
six panels with MW, as in the image below. If you don't have ATSAS |
| 64 |
installed you will see just four panels, the rightmost panel in each |
| 65 |
row will be missing. |
| 66 |
|
| 67 |
* *Tip:* To learn more about any of the methods, click on the "More Info" |
| 68 |
button. |
| 69 |
|
| 70 |
|mw_png| |
| 71 |
|
| 72 |
#. In either concentration box, enter the sample concentration of 0.47 mg/ml. Notice that you |
| 73 |
now get results from all methods of MW calculation. |
| 74 |
|
| 75 |
* *Question:* The expected MW value for GI is 172 kDa. How do your results compare? |
| 76 |
|
| 77 |
#. Click on the “Show Details” button for the Vc MW panel. You should see a graph, which shows |
| 78 |
the integrated area of :math:`qI(q)` vs. *q*\ . For this method to be accurate, |
| 79 |
this value needs to converge at high *q*\ . |
| 80 |
|
| 81 |
|mw_vc_png| |
| 82 |
|
| 83 |
#. Click on the "Show Details" for the Vp MW panel. You'll notice that both Vp MW |
| 84 |
and Vc MW have a "cutoff" selection. At higher q you can start to get |
| 85 |
scattering from flexibility or intra-molecular features that may reduce the |
| 86 |
reliability of the MW estimate. RAW automatically cuts off the scattering profile |
| 87 |
used based on the size of the object. You can change the cutoff if you need to. |
| 88 |
|
| 89 |
|mw_cutoff_png| |
| 90 |
|
| 91 |
#. Click the “OK” button to save your analysis. |
| 92 |
|
| 93 |
* *Note:* The “Cancel” button discards the analysis. |
| 94 |
|
| 95 |
* *Tip:* After clicking "OK" you can now click on the GI profile in the |
| 96 |
Profiles control panel and see the MW you just found in the Info panel. |
| 97 |
|
| 98 |
#. Repeat the MW analysis for the lysozyme sample, which had a concentration of 4.27 mg/ml. |
| 99 |
The expected MW of lysozyme is 14.3 kDa. |
| 100 |
|
| 101 |
* *Question:* Does the Vc method work for the lysozyme data? |
| 102 |
|
| 103 |
|
| 104 |
.. |mw_open_png| image:: images/mw_open.png |
| 105 |
:width: 400 px |
| 106 |
:target: ../_images/mw_open.png |
| 107 |
|
| 108 |
.. |mw_png| image:: images/mw.png |
| 109 |
:target: ../_images/mw.png |
| 110 |
|
| 111 |
.. |mw_vc_png| image:: images/mw_vc.png |
| 112 |
:width: 300 px |
| 113 |
:target: ../_images/mw_vc.png |
| 114 |
|
| 115 |
.. |mw_cutoff_png| image:: images/mw_cutoff.png |
| 116 |
:width: 600 px |
| 117 |
:target: ../_images/mw_cutoff.png |