# HydraSDR Calibration Procedure ## Overview This procedure allows you to calibrate the frequency accuracy of your HydraSDR software-defined radio device by measuring and correcting frequency errors using parts-per-billion (PPB) corrections. ## Step-by-Step Calibration 1. **Clear existing calibration**: Reset any previous calibration by writing zero correction: ```bash hydrasdr_calibrate -w 0 ``` 2. **Restart the device**: Power cycle the HydraSDR software-defined radio or use `hydrasdr_reset` to ensure the new calibration is loaded during boot initialization 3. **Measure frequency error**: Use your measurement tool to compare actual frequency output against expected frequency, then calculate the correction needed using a PPM/PPB calculator (such as https://www.sitime.com/ppm-hz-calculator) - round the result to the nearest PPB value 4. **Apply the calculated correction**: Write the computed PPB value using `hydrasdr_calibrate -w `: ```bash hydrasdr_calibrate -w -1942 ``` 5. **Restart the device again**: Power cycle the HydraSDR software-defined radio or use `hydrasdr_reset` to load and apply the new calibration settings 6. **Verify calibration**: Test with your measurement software to confirm the frequency correction is now accurate ## Important Notes - Device restart is required after each calibration change because the calibration data is only read during the boot process - Both positive and negative PPB correction values are supported - The calibration data includes a timestamp and is stored in flash memory sector 2 - Use `hydrasdr_calibrate -r` to read current calibration values ## Example If your measurement shows the device is running 1942 PPB slow, you would apply a positive correction: ```bash # Clear existing calibration hydrasdr_calibrate -w 0 # Restart device hydrasdr_reset # After measurement and calculation hydrasdr_calibrate -w 1942 # Restart device to apply hydrasdr_reset # Verify with your measurement tool ```